CHAPTER
Saf-C 6300 BREATH
ALCOHOL PROCEDURE USING THE INTOXILYZER 5000
INSTRUMENT
BEFORE JANUARY 1, 2017 – See Saf-C 7100 for use on or after January 1,
2017
REVISION NOTE:
Pursuant to 2003, 319:80, II, effective 1-1-04, rules He-P
2207 entitled “Breath Alcohol Procedure Using the Intoxilyzer
5000 Instrument”, last filed by the Department of Health and Human Services
under Document #6208, effective 4-1-96, were transferred to the Department of
Safety. The Department of Safety
subsequently filed Document #8064, effective 4-1-04, which readopted and
renumbered as interim rules in Saf-C 6300 most of the
rules in the former He-P 2207.
The following rules in He-P 2207 were not included within
Document #8064 and are deemed by the Department of Safety to have expired on
4-1-04:
Definitions in He-P
2207.01(q) “Commissioner”; (s) “Division”, (w) “Forensic breath test supervisor
II”; (ab) “Laboratory”; and (ag) “Proficiency testing program”, and
He-P 2201.18 relative to a
proficiency testing program.
The rules in Document #8064 therefore replaced all prior
filings for rules in the former He-P 2207.
The filings of the Department of Health and Human Services or its Division
of Public Health Services affecting the former He-P 2207 include the following
documents:
#4744-a, effective 1-24-90
#6145, INTERIM, effective
12-21-95
#6208, effective 4-1-96
PART Saf-C 6301 SCOPE
Saf-C 6301.01 Applicability. This chapter shall apply to all breath
alcohol procedures using the intoxilyzer 5000
instrument for breath tests conducted before January 1, 2017.
Source. #12064, INTERIM, eff 1-1-17, EXPIRES:
6-30-17; ss by #12216, eff 6-30-17
PART Saf-C
6302 DEFINITIONS
Saf-C 6302.01
“Accuracy” means the closeness of the test result to the true or
expected result.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04 (See
Revision Note at Chapter Heading for Saf-C 6300); ss by #8114, eff 7-14-04; ss by
#9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6301.01)
Saf-C 6302.02
“Acetone” means dimethylketone, a volatile
chemical which can be found in the blood, urine or breath of a person under
certain metabolic conditions such as diabetes.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly
Saf-C 6301.02)
Saf-C 6302.03
“Alcohol” means ethanol, ethyl alcohol or “intoxicating liquors” as
defined in RSA 21:33.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly
Saf-C 6301.03)
Saf-C 6302.04
“Alcohol concentration” means grams of alcohol per 210 liters of breath
(g/210L) as defined in RSA 259:3-b.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C 6301.04)
Saf-C 6302.05
“Alveolar air” means that portion of a breath sample submitted which is
nearest in exchange between the lungs and the blood, the last portion of a
maximal expiration.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.04);
renumbered by #12064 (formerly Saf-C 6301.05)
Saf-C 6302.06
“Approval” means a statement of authorization signed by the
commissioner.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04 (See
Revision Note at Chapter Heading for Saf-C 6300); ss by #8114, eff 7-14-04; ss by
#9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6301.06)
Saf-C 6302.07
“Approved course of training” means a course of instruction approved by
the commissioner for proper operation of the breath test instrument.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6301.07)
Saf-C 6302.08
“Approved instrument” means the Intoxilyzer
5000.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6301.08)
Saf-C 6302.09
“Breath test operator” means an individual who has completed the
eligibility requirements in Saf-C 6303.02.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.11); renumbered
by #12064 (formerly Saf-C 6301.09)
Saf-C 6302.10
“Captured sample” means that sample of breath introduced into an
approved instrument and adsorbed onto a silica gel tube.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6301.10)
Saf-C 6302.11 “Captured
sample bag” means a pre-numbered tamper resistant envelope in which the
subject’s captured samples are placed after completion of the breath test.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.13);
renumbered by #12064 (formerly Saf-C 6301.11)
Saf-C 6302.12
“Certification of breath test operator” means a 12 month period in which
a law enforcement officer, who meets requirements of Saf-C
6303.02, is authorized to conduct an analysis of a person’s breath for the
purpose of determining its alcohol concentration.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.14);
renumbered by #12064 (formerly Saf-C 6301.12)
Saf-C 6301.13
“Certification of equipment” means a 6 month interval in which the
accuracy and precision of a breath test instrument have been verified by the forensic breath
testing supervisor.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.15);
renumbered by #12064 (formerly Saf-C 6301.13)
Saf-C 6302.14
“Commissioner” means the commissioner of the department of safety or his
or her designee.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.17);
renumbered by #12064 (formerly Saf-C 6301.14)
Saf-C 6302.15
“Control” means a sample or solution of a known concentration, similar
to the sample which is analyzed, for quality control purposes and not used for
quality control purposes and not used in the calibration of a method.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.18);
renumbered by #12064 (formerly Saf-C 6301.15)
Saf-C 6302.16
“External standard reference” means a simulator test performed on the
approved instrument, using a known alcohol/water solution, to verify the accuracy
of the approved instrument.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.20);
renumbered by #12064 (formerly Saf-C 6301.16)
Saf-C 6302.17
“Field” means a police department or law enforcement agency located
within
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.21);
renumbered by #12064 (formerly Saf-C 6301.17)
Saf-C 6302.18
“Forensic breath test supervisor” means an individual who is authorized
by the commissioner and has met the requirements of Saf-C
6303.01.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.22);
renumbered by #12064 (formerly Saf-C 6301.18)
Saf-C 6302.19 “Gas
chromatography” means a method of analysis in which a sample is injected into a
stream of carrier gas moving through a column containing a stationary phase
composed of a liquid or particulate solid.
The sample is then separated into its component compounds according to
their affinity for the stationary phase.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.23);
renumbered by #12064 (formerly Saf-C 6301.19)
Saf-C 6302.20
“Headspace vapor” means a volume of gaseous sample above a liquid in a
closed container.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C 6300);
ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.24);
renumbered by #12064 (formerly Saf-C 6301.20)
Saf-C 6301.21
“Internal standard” means a substance which is added to the sample vial
in a known concentration, which is chemically similar to the analyte being determined and against which the analyte can be measured.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.25);
renumbered by #12064 (formerly Saf-C 6301.21)
Saf-C 6302.22
“Internal standard reference” means an automatic check conducted by the
approved instrument to determine the readiness of that instrument to analyze a
sample of breath.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.26);
renumbered by #12064 (formerly Saf-C 6301.22)
Saf-C 6302.23
“Laboratory” means the forensic science laboratory or a laboratory
accredited pursuant to RSA 265-A:5,V.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.27);
renumbered by #12064 (formerly Saf-C 6301.23)
Saf-C 6302.24
“Modification” means an alteration to the approved instrument affecting
the accuracy or precision of the breath test.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.29);
renumbered by #12064 (formerly Saf-C 6301.24)
Saf-C 6302.25
“Precision” means a measure of the agreement between replicate tests.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.30);
renumbered by #12064 (formerly Saf-C 6301.25)
Saf-C 6302.26
“Preventive maintenance check” means a scheduled or routine maintenance
performed on the approved instrument to ensure its continued accuracy and
precision.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.31);
renumbered by #12064 (formerly Saf-C 6301.26)
Saf-C 6302.27
“Recertifying supervisor” means a breath test operator, authorized by
the commissioner, who has met the eligibility requirements in Saf-C 6303.02.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.33);
renumbered by #12064 (formerly Saf-C 6301.27)
Saf-C 6302.28
“Repairs” means an adjustment or replacement of parts or components to
return the equipment to proper operating condition.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.34);
renumbered by #12064 (formerly Saf-C 6301.28)
Saf-C 6302.29
“Simulator” means a device used as an external standard reference which
is designed to produce a similar result as that obtained from an actual breath
sample, when operated under predefined conditions.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.35);
renumbered by #12064 (formerly Saf-C 6301.29)
Saf-C 6302.30
“Standard” means a silica gel tube containing a known concentration of analyte.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.36);
renumbered by #12064 (from Saf-C 6301.30)
Saf-C 6302.31
“Subject sample” means a sample of alveolar air, given by a person being
tested to determine a breath alcohol concentration.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6301.37);
renumbered by #12064 (formerly Saf-C 6301.31)
PART Saf-C 6303 EQUIPMENT
Saf-C 6303.01 Approval
of Equipment.
(a) The commissioner shall approve the breath
test instrument and simulator by make and model in accordance with (b)
below. Pursuant to RSA 265-A:12, breath tests conducted on an
approved instrument to determine the alcohol concentration of one’s breath
shall be offered for evidentiary purposes when performed by certified personnel
in accordance with procedures established in Saf-C
6302.02.
(b) The following criteria shall be used by the
commissioner for approval of the breath test instrument:
(1) Breath
samples collected for analysis shall be alveolar in composition;
(2) The instrument
shall be capable of analyzing a series of headspace vapor samples from a
simulator at known concentrations;
(3) The
instrument shall be capable of performing an internal standard reference check;
(4) The
instrument shall be capable of capturing the subject’s breath sample for
delayed analysis;
(5) The
instrument shall be identified on the “Conforming Products List” by the
National Highway Traffic Safety Administration as an “Evidential Breath
Measurement Device.”
(c) Any approved instrument submitted to the
manufacturer for modification or repair shall be returned to the forensic
science laboratory upon completion of servicing and shall pass a preventive
maintenance check in accordance with Saf-C 6302.04
prior to returning the approved instrument to the field.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6302.01)
Saf-C 6303.02 Sample
Collection.
(a) All breath samples shall be alveolar air and
shall be collected at the direction of the breath test operator.
(b) Immediately prior to testing, the subject shall
be observed by a breath test operator for a period of 20 minutes to ensure the
integrity of the sample. Prior to the
observation period, the breath test operator shall inform the subject not to
place any material in his/her mouth.
Should vomiting, regurgitation or belching occur or any external
material enter the mouth of the person to be examined, an additional 20 minute
wait period shall occur before testing.
Should the breath test operator determine that the subject is not
complying with the requirements of this paragraph or (d) below at any point
during the testing process, or the subject decides not to complete the sampling
requirements in (d), the breath test operator shall stop the test and record
the test as a refusal pursuant RSA 265-A:8.
(c) The breath test operator shall record and
shall print out on the test record the identification of the subject, the name
of the breath test operator, and the date and time of the sample collection. Identification of the approved instrument
used to conduct the test shall include the make, model, serial number and field
location.
(d) A minimum of 2 subject samples shall be
collected for each complete breath test.
If the alcohol concentration readings of the first 2 samples do not
agree within +/- 0.02 g/210L, a third subject sample shall be collected.
(e) A reported value shall be printed on the test
record upon completion of the following:
(1) A 20 minute
observation period before the initial sample is taken;
(2) The
identification of the subject, breath test operator and instrument;
(3) An ambient
or room air check performed by the approved instrument;
(4) An internal
standard reference check performed by the approved instrument;
(5) The
satisfactory collection of a minimum of 2 subject samples which agree with +/-
0.02 g/210L;
(6) The
collection of a captured sample for each satisfactory subject sample collected
in accordance with (a), (b) and (d);
(7) An external
standard reference check performed by the approved instrument in accordance
with (h) below; and
(8) The
printout of test results and required information.
(f) During the testing process there shall be 2
reference checks on the approved instrument.
(g) The approved instrument shall be programmed to
perform an internal standard reference check at separate set points. If any one of the set points gives a reading
greater than +/- 5% of the assigned value, the approved instrument shall be so
designed that it shall not allow a subject sample to be taken.
(h) An external standard reference check shall be
conducted via the external reference port on the approved instrument. The external standard reference check shall
consist of a known alcohol/water solution held at a constant temperature of
34°C +/- 0.2°C, whose vapor shall give an alcohol concentration reading on the
approved instrument of no less than 0.09 g/210L nor greater than 0.10
g/210L. The approved instrument shall be
so designed that a test cannot be completed if the external standard reference
result is not within the acceptable range.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6302.02)
Saf-C 6303.03 Reporting
of Findings.
(a) All completed test results shall be printed
on the breath test record and shall be reported to the second decimal place
without rounding.
(b) A reported value shall be printed when a
breath test is completed in accordance with Saf-C
6302.02(e) above;
(c) The reported value shall be the lower of the
2 values which are within +/- 0.02 g/210L agreement.
(d) Breath test results shall be expressed as
grams of alcohol per 210L.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6302.03)
Saf-C 6303.04 Preventive
Maintenance-Equipment.
(a) The forensic breath testing supervisor shall
conduct a preventive maintenance on each approved instrument 6 months after the
initial check and at 6 month intervals.
(b) A preventive maintenance check pursuant to
RSA 265-A:12,II shall include:
(1) Analysis of
a series of simulated breath tests at known alcohol concentrations, to verify
the accuracy and precision of the approved instrument which shall meet the
following criteria:
a. For known
alcohol concentrations less than 0.10g/210L:
1. The accuracy
of the readings shall deviate no more than +/- 0.005 g/210L from the known
alcohol concentration; and
2. The total precision
range of the readings shall not exceed +/- 0.005 g/210L; and
b. For known
alcohol concentrations equal to or greater than 0.10 g/210L:
1. The accuracy
of the readings shall be within +/- 5% of the known alcohol concentration; and
2. The total
precision range of the readings shall not exceed 5% of the target value;
(2)
Verification of the approved instrument’s ability to detect and correct
for acetone;
(3) Collection
of captured samples at known alcohol concentrations to verify the accuracy of
the captured sample method which shall meet the following criteria:
a. For infrared
values less than 0.10 g/210L, results shall agree within 0.01 g/210L of the
infrared value; and
b. For infrared
values equal to or greater than 0.10 g/210L, results shall agree within +/- 10%
of the infrared value;
(4) Performance
of routine maintenance on the approved instrument and simulator as necessary to
maintain proper analytical functions; and
(5)
Documentation of maintenance performed on each approved instrument and
simulator at the time of service.
(c) The forensic breath testing supervisor shall
certify the accuracy and precision of the approved instrument on the preventive
maintenance check form, pursuant to RSA 265-A:12, II, once the requirements of
(b)(1) – (5) are completed.
(d) The preventive maintenance check form shall
specify the following:
(1) The
approved instrument and simulator by make, model and serial number;
(2) The field
location of the equipment;
(3) The date of
certification and expiration; and
(4) That the forensic breath testing supervisor has
found the approved instrument and simulator to be in proper working order and
that the equipment meets the preventive maintenance requirements as outlined in
(b) above and RSA 265-A:12, II.
(e) Repairs
performed on the approved instrument shall be documented and kept on file at
the forensic science laboratory. The
approved instrument shall not be required to pass a preventive maintenance
check if the repair does not affect the accuracy or precision of the breath
test.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6302.04)
PART Saf-C 6304 QUALIFICATIONS OF INDIVIDUALS CONDUCTING
TESTS
Saf-C 6304.01 Forensic
Breath Testing Supervisor.
(a) Pursuant to RSA 265-A:5, the commissioner
shall designate a forensic breath testing supervisor as competent and qualified
to conduct such breath tests.
(b) To be considered for designations as a
forensic breath test supervisor, an applicant shall meet the following
requirements:
(1) A forensic
breath testing supervisor shall meet the minimum qualifications under division
of personnel position classification requirements as a laboratory scientist, a
criminalist or a toxicologist; and
(2) Complete
the manufacturer’s equipment maintenance course for the approved instrument.
(c) Upon designation by the commissioner, the
forensic breath testing supervisor shall perform the following duties:
(1) Perform
preventive maintenance on the approved instrument;
(2) Certify the
accuracy and precision of the approved instrument as required in Saf-C 6302.04;
(3) Train and
recertify breath test operators and recertifying supervisors; and
(4) Evaluate
new breath test instruments, as necessary, following the criteria for approval
in accordance with Saf-C 6302.01(b).
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6303.01)
Saf-C 6304.02 Breath
Test Instrument Operator/Recertifying Supervisor.
(a) Personnel certified by the commissioner, pursuant
to RSA 265-A:5, V, shall be designated as a breath test instrument operator or
as a recertifying supervisor.
(b) Criteria for breath test operators and
recertifying supervisors shall be as follows:
(1) To be
eligible for training as a breath test operator of the approved instrument, for
purposes of RSA 265-A:4, the applicant shall be currently employed as a law
enforcement official within the state of
(2) To be
eligible as a recertifying supervisor:
a. The
applicant shall meet the eligibility requirements in (1) above;
b. The
applicant shall have taken and passed an approved course of training as
outlined in (h) below;
c. The
applicant shall be currently certified as a breath test instrument operator;
d. The applicant
shall have experience in conducting actual breath tests by:
1. Maintaining
the certification in accordance with (k) below for a period of not less than
one year; and
2. Conducting a
minimum of 10 subject tests in accordance with Saf-C 6302.02
during one calendar year.
(c) A breath test operator who has met the
eligibility requirements of (b) above and has passed the approved course of
training as outlined in (h) below shall be approved by the commissioner to
conduct breath tests pursuant to RSA 265-A:5.
(d) A recertifying supervisor who has met the
eligibility requirements of (b)(2) above shall perform the following duties:
(1) Conduct
breath tests pursuant to RSA 265-A:5;
(2) Recertify
breath test operators within their own agency in accordance with (j) below; and
(3) Assist the
forensic breath testing supervisor in the training of breath test operators.
(e) Recertifying supervisors shall not recertify
equipment.
(f) Applicants shall meet the eligibility requirements
in (b) above and have completed a course of training provided by the
division covering the topics outlined in (h) below. A certificate shall be issued to the breath
test operator by the agency sponsoring the course of training. This certificate of training shall be
submitted to the commissioner when applying for certification as a breath test
operator.
(g) A comprehensive practical and written
examination shall be developed and administered by the forensic breath testing
supervisor at the completion of the course.
(h) The exam required in accordance with (g)
shall cover the following subject matter:
(1) The value and purpose of breath alcohol
testing;
(2) The
physiological action of alcohol on the human body;
(3) The pharmacology
and toxicology of alcohol;
(4) The methods
of alcohol analysis and theory of breath testing;
(5) The instruments and procedures for breath
testing;
(6) Laboratory
practice and demonstration of competency by:
a. Using proper
technique in collecting and capturing a sample;
b. Obtaining
accurate results; and
c. Following
established operating and reporting procedures; and
(7) Legal
procedures and decisions affecting breath testing.
(i) A breath test
operator or recertifying supervisor certification shall be valid for a period
of 12 months from date of issue unless terminated or revoked.
(j) A breath test operator shall renew his/her
certification prior to its expiration through:
(1) Demonstration
of competence in performing an analysis under the supervision of a forensic breath testing supervisor or
recertifying supervisor by:
a. Using proper technique in collecting and
capturing a sample;
b. Obtaining
accurate results; and
c. Following
established operating and reporting procedures; and
(2) Completion
of a review course provided by the
division which shall include:
a. Instrument
operating and sampling technique;
b. Analysis and
reporting procedures; and
c. Procedural
changes and legal decisions affecting breath testing.
(k) A certified recertifying supervisor shall
renew his/her certification prior to its expiration through:
(1)
Demonstration of competence in performing an analysis under the supervision
of a forensic breath testing supervisor by:
a. Using proper technique in collecting a
sample;
b. Obtaining
accurate results; and
c. Following
established operating and reporting
procedures; and
(2) Completion
of a review course provided by the division which shall include:
a. Instrument
operating and sampling technique;
b. Analysis and
reporting procedures; and
c. Procedural
changes and legal decisions affecting breath testing.
(l) Reinstatement of breath test operators and recertifying
supervisors shall be as follows:
(1) A breath
test operator's certification shall be reinstated by the forensic breath
testing supervisor or recertifying supervisor if he or she meets the
requirements of (j) above within one year from the date of expiration; and
(2) A
recertifying supervisor's certification shall be reinstated by the forensic
breath testing supervisor if he or she meets the requirements of (k) above
within one year from the date of expiration.
(m) A breath test operator's or recertifying
supervisor's certification shall remain valid unless expired or revoked after
notice and opportunity for a hearing for
the following reasons:
(1) Failure to
follow the operating procedure for
the approved instrument in accordance with Saf-C
6302.02(e);
(2) Failure to
follow the procedure for reporting and keeping of records;
(3) Failure to
renew certification within a 12 month period prior to expiration;
(4) Termination
of employment as a law enforcement official at any time during the 12 month
certification period, except for those individuals transferring from one law
enforcement agency to another within the State of New Hampshire; and
(5) Failure to
become reinstated after a period of one year from expiration of their last
certification.
(n) To become certified again, the former breath
test operator or recertifying supervisor shall attend and complete an approved
course of training as described in (h) above.
(o) A
breath test operator or recertifying supervisor's certification shall be
revoked upon recommendation by the forensic breath testing supervisor in
accordance with (m)(1) and (2) above.
Notification of this revocation shall be forwarded to the breath test
operator or recertifying supervisor by the commissioner by certified mail.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6303.02)
PART Saf-C 6305 BREATH ALCOHOL SAMPLES
Saf-C 6305.01 Preservation
of Breath Alcohol Sample.
(a) When a subject submits to a breath test on an
approved instrument to determine his/her alcohol concentration, pursuant to RSA
265-A:4 a captured sample of their breath shall be collected by the breath test
operator.
(b) The breath test operator shall identify each
captured sample of breath collected in accordance with Saf-C
6302.02 by the following:
(1) The lot
number of the tube;
(2) The name of
the individual tested;
(3) The date of
collection; and
(4) The number
in the sequence of samples collected.
(c) The captured samples of breath shall be
placed in a captured sample bag along with a copy of the subject's test record.
(d) The breath test operator shall record on the outside of the captured sample bag:
(1) The
subject's name; and
(2) The date
the captured samples were collected.
(e) Each subject who submits to a breath test and
successfully completes the requirements of Saf-C
6302.02(e) shall be given the captured samples of his or her breath at the
completion of the test.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6304.01)
PART Saf-C 6306 INDEPENDENT LABORATORY ACCREDITATION
Saf-C 6306.01 Laboratory
Accreditation, Independent Analysis. Any laboratory that performs an
analysis of a captured sample to determine its alcohol concentration pursuant
to RSA 265-A:7, shall be accredited by a nationally recognized accrediting body
pursuant to RSA 265-A:5, V.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10; renumbered by #12064 (formerly Saf-C
6305.01)
Saf-C 6306.02 Method
of Analysis of Captured Samples-Independent Laboratory.
(a) An independent laboratory accredited pursuant
to RSA 265-A:5, V shall employ the following in its analytical method:
(1) Gas
chromatography;
(2) Internal
standard;
(3) Captured
sample standards prepared with the same type and amount of trapping material as
utilized to collect a captured sample of breath;
(4) Captured
sample standards prepared on the same series of approved instrument as used in
the original collecting procedure;
(5) Analysis of
the captured sample which corresponds to the reported value as determined in Saf-C 6302.03; and
(6) Analysis of
the captured sample identified in (5) above in duplicate.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6305.04);
renumbered by #12064 (formerly Saf-C 6305.02)
Saf-C 6306.03 Reporting
of Findings.
(a) The reported value of a captured sample
analyzed in accordance with Saf-C 6305.02 shall be
determined by:
(1) Taking the
lower of the 2 duplicate results;
and
(2) Truncating
the result determined in (1) to the second decimal place.
(b) The reported value determined in (a) shall be
expressed as grams of alcohol per 210 liters of breath (g/210L), pursuant to
RSA 259:3-b.
(c) The deviation between the reported value of
the approved instrument and the independent laboratory’s reported value shall
be derived by subtracting the independent laboratory’s reported value
determined in (a) from the reported value of the approved instrument.
(d) The deviation determined in (c) shall be
expressed as grams of alcohol per 210 liters of breath (g/210L).
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04 (See
Revision Note at Chapter Heading for Saf-C 6300); ss by #8114, eff 7-14-04; ss by
#9683, eff 3-23-10 (from Saf-C 6305.07); renumbered
by #12064 (formerly Saf-C 6305.03)
Saf-C 6306.04 Captured
Sample Analysis Report.
(a) The report of the analysis of a captured
sample performed by an independent laboratory pursuant to RSA 265-A:7 shall
include:
(1) The serial
number of the captured sample bag;
(2) The serial
number of the approved instrument from which the breath sample was captured;
(3) The
subject’s reported value identified on the printout from the approved
instrument in accordance with Saf-C 6302.04;
(4) The results
of the captured sample analyzed in duplicate in accordance with Saf-C 6305.02(a)(5) and (6);
(5) The
independent laboratory’s reported value as determined in accordance with Saf-C 6305.03(a); and
(6) The name of
the person conducting the analysis.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6305.08);
renumbered by #12064 (formerly Saf-C 6305.04)
Saf-C 6306.05 Interpretation
of Results.
(a) Captured samples of breath collected in accordance
with Saf-C 6302.02(e) and Saf-C
6304.01 and analyzed by an independent laboratory in accordance with Saf-C 6305.02-04 shall be used to corroborate the reported
value obtained by the approved instrument.
(b) The result of the breath test shall be considered
to be corroborated when the reported value of the approved instrument is within
the deviation limits for the alcohol concentration levels outlined below:
(1) For alcohol
concentrations less than 0.05 g/210L and the deviation does not exceed 0.01
g/210L;
(2) For alcohol
concentrations equal to 0.05 g/210L but less than 0.10 g/210L and the deviation
does not exceed 0.02 g/210L;
(3) For alcohol
concentrations equal to 0.10 g/210L but less than 0.15 g/210L and the deviation
does not exceed 0.03 g/210L;
(4) For alcohol
concentrations equal to 0.15 g/210L but less than 0.20 g/210L and the deviation
does not exceed 0.04 g/210L; and
(5) For alcohol
concentrations equal to or greater than 0.20 g/210L and the deviation does not
exceed 0.05 g/210L.
(c) When the result of the breath test is not
corroborated in (b) above, further evaluation shall be required in order to
determine the weight to be given to the reported value of either the approved
instrument or the independent laboratory.
(d) Pursuant to RSA 265-A:12, the evaluation of
results shall include a review of the following materials provided by:
(1) The
arresting officer or breath test operator to include:
a. A copy of
the preventive maintenance check form completed in accordance with Saf-C 6302.04 and dated prior to the subject’s test;
b. A copy of
the documentation of any repairs performed on the approved instrument within
the certification period encompassing the subject’s test; and
c. A copy of
the breath test printout showing the completed test, in accordance with Saf-C 6302.03; and
(2) The
independent laboratory to include:
a. Proof of the
independent laboratory’s accreditation as required by RSA 265-A:5, V(e) for the
period encompassing the date of analysis;
b. A copy of
the captured sample analysis report;
c. A copy of
the run log for the day the subject’s captured sample was analyzed;
d. Copies of
the gas chromatograph instrument printouts of the subject’s captured sample
results which correspond to the reported value obtained by the approved
instrument;
e. Copies of
all the gas chromatograph instrument printouts of the blanks, standards and
controls analyzed on the day the captured sample was analyzed; and
f.
Documentation of any repairs or preventive maintenance performed on the
gas chromatograph instrument within a 2 month period before and after the
analysis of the subject’s captured sample.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6305.09);
renumbered by #12064 (formerly Saf-C 6305.05)
Saf-C 6306.06 Statistics.
(a) Laboratories accredited to perform
independent analyses of captured samples of breath shall submit a monthly
summary of their results to the forensic science laboratory. The summary shall be submitted no later than
10 working days after the end of
each month.
(b) The summary shall consist of:
(1) The
approved instrument serial number;
(2) The date of
the breath test and the reported value; and
(3) The
independent laboratory’s reported value determined in accordance with Saf-C 6305.03.
(c) The summary information shall be submitted in
one of the following formats:
(1) Written on
8.5 inch by 11 inch paper;
(2) Via e-mail
with an attached Microsoft Excel, Word or ASCII file.
(d) The summary data shall be used by the forensic science laboratory to monitor the
performance of the approved instruments to ensure the accuracy of the testing
program.
Source. #8064, INTERIM, eff 4-1-04, EXPIRES: 9-28-04
(See Revision Note at Chapter Heading for Saf-C
6300); ss by #8114, eff 7-14-04; ss
by #9683, eff 3-23-10 (from Saf-C 6305.10);
renumbered by #12064 (formerly Saf-C 6305.06)
APPENDIX
Rule |
|
|
|
Saf-C 6301 |
RSA 541-A:7 |
Saf-C 6302 |
RSA 541-A:7 |
Saf-C 6303.01(a) |
RSA 265-A:5, V & RSA 265-A:12, IV |
Saf-C 6303.01(b) to 6303.03 |
RSA 265:-A:5, V |
Saf-C 6303.04(a)–(d) |
RSA 265-A:V, V(f) & RSA 265-A:12, II |
Saf-C 6303.04(e) |
RSA 265-A:5, V(f) |
Saf-C
6304 |
RSA 265:85,
V(b) |
Saf-C
6305 |
RSA 265:85,
V(c) & RSA 265:86, II |
Saf-C
6306.01 |
RSA 265:85,
V(e) |
Saf-C 6306.02-6306.04 |
RSA 265-A:5, V(a) & (e) |
Saf-C 6306.05(a)-(c) |
RSA 265-A:5, V(a) &(e) |
Saf-C 6306.05(d)(1)a. |
RSA 265-A:5, V(d) & RSA 265-A:12, II |
Saf-C 6306.05(d)(1)b.–6306.05(d)(2) |
RSA 265-A:5, V(d) |
Saf-C 6306.06 |
RSA 265-A:5, V(e) &(f) |