CHAPTER Saf-C 6800  PRELIMINARY BREATH TEST DEVICES

 

Statutory Authority:  RSA 265-A: 15, III

 

REVISION NOTE:

 

          Pursuant to 2003, 319:80, II, effective 1-1-04, rules He-P 2206 entitled “Preliminary Breath Test Devices”, last filed by the Department of Health and Human Services under Document #7781, effective 10-23-02, were transferred to the Department of Safety.  The Department of Safety subsequently filed Document #9685, effective 3-23-10, which readopted with amendments and renumbered He-P 2206 as Saf-C 6800.

 

          The rules in Document #9685 replaced all prior filings for rules in the former He-P 2206.  The filings of the Department of Health and Human Services or its Division of Public Health Services affecting the former He-P 2206 include the following documents:

 

                  #2309, eff 1-1-83

                  #2340, eff 4-1-83

                  #2476, eff 9-14-83

                  #3068, eff 7-23-85

                  #4044, eff 4-25-86, EXPIRED 4-25-92

                  #7781, eff 10-23-02

 

PART Saf-C 6801  DEFINITIONS

 

          Saf-C 6801.01 “Alcohol” means ethanol, ethyl alcohol, or “intoxicating liquors” as defined in RSA 21:33.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6801.02 “Approval” means a statement of authorization signed by the commissioner.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6801.03 “Commissioner” means the commissioner of the department of safety or his or her designee.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6801.04 “Laboratory” means the New Hampshire department of safety, division of state police, forensic laboratory.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

PART Saf-C 6802  APPROVAL OF PRELIMINARY BREATH TEST DEVICES

 

          Saf-C 6802.01  Criteria for Approval.

 

          (a)  Pursuant to RSA 265-A:15,II, the commissioner shall approve the preliminary breath test device by manufacturer make and model in accordance with (b) below.

 

          (b)  The criteria to be used for approval of a preliminary breath test device shall be the following:

 

(1) The device shall be identified on the National Highway Traffic Safety Administration’s “Conforming Products List” of evidential breath measurement devices;

 

(2) The device shall be capable of analyzing a series of samples using a simulator or dry gas standard at known alcohol concentrations; and

 

(3)  The device, submitted in accordance with Saf-C 6802.02, shall successfully pass the protocol for evaluation of preliminary breath test devices in accordance with Saf-C 6802.03.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6802.02  Manufacturer Responsibility.

 

          (a)  A manufacturer of preliminary breath test devices may submit a request for an evaluation of the device by the laboratory.

 

          (b)  Each device submitted for evaluation shall be accompanied by any peripheral materials necessary to perform the evaluation such as:

 

(1)  An operation manual;

 

(2)  A list of unit capabilities that include criteria identified in Saf-C 6802.03 (b);

 

(3)  The carrying case;

 

(4)  Mouthpieces;

 

(5)  Batteries; and

 

(6)  Peripheral hardware and software.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6802.03  Laboratory Evaluation.

 

          (a)  The laboratory shall evaluate each manufacturer make and model of preliminary breath test device submitted in accordance with Saf-C 6802.02.

 

          (b)  The evaluation shall include the following categories:

 

(1)  Analysis of a series of alcohol samples at known concentrations, to verify the accuracy and precision of the approved instrument as meeting the following criteria:

 

a.  For known alcohol concentrations greater than 0.000 g/210L but less than or equal to

0.050 g/210L:

 

1.  The accuracy of the readings shall deviate no more than +/- 0.003 g/210L from known alcohol concentrations; and

 

2.  The total precision range of the readings shall not exceed 0.005 g/210L;

 

b.  For known alcohol concentrations greater than 0.050 g/210L but less than or equal to 0.100 g/210L:

 

1.  The accuracy of the readings shall deviate no more than +/- 0.005 g/210L from known concentrations; and

 

2.  The total precision range of the readings shall not exceed 0.005 g/210L; and

 

c.  For known alcohol concentrations greater than 0.100 g/210L:

 

1. The accuracy of the readings shall deviate no more than +/- 5% from known alcohol concentrations; and

 

2.  The total precision range of the readings shall not exceed 5% of the target value;

 

(2)  Ease of use in handling the device while performing a test;

 

(3)  Performance under normal and adverse ambient conditions; and

 

(4) Effects of other substances or sources of energy that might interfere with the proper functioning of the device.

 

          (c)  Each device evaluated in accordance with Saf-C 6802.03 shall meet or exceed the performance statements provided by the manufacturer.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

                 Saf-C 6802.04  Reporting of Findings.

 

          (a) A written report shall be made and submitted to the commissioner upon completion of the evaluation.

 

          (b)  The report shall include an evaluation of all categories listed in Saf-C 6802.02 including data collected from each device.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6802.05  Approval.

 

          (a)  A letter of approval shall be issued from the commissioner to an equipment manufacturer whose preliminary breath test device has met the requirements of Saf-C 6802.01 and Saf-C 6802.02. 

 

          (b)  Any device previously approved by the commissioner prior to adoption of Saf-C 6800 shall be considered valid for purposes of RSA 265-A:15.

 

          (c)  The commissioner shall maintain a list of preliminary breath test devices approved for use in New Hampshire.  This list shall be available to anyone upon request.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6802.06  Denial.

 

          (a)  The commissioner shall deny the approval of any preliminary breath test device that fails to meet the criteria for approval as outlined in accordance with Saf-C 6802.01.

 

          (b)  The commissioner shall notify the manufacturer of the denial in writing outlining the areas of noncompliance.

 

          (c)  Any manufacturer whose preliminary breath test device was not approved by the commissioner in accordance with Saf-C 6802.05 may appeal the decision and request a hearing in accordance with Saf-C 203.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 

          Saf-C 6802.07  Resubmission.

 

          (a)  An equipment manufacturer whose request for approval of preliminary breath test device has been denied in accordance with Saf-C 6802.06 may resubmit the device for reconsideration of evaluation after a 6 month period from the date of the letter of denial. The device submitted for reconsideration shall be evaluated only if there has been improvements or modifications to the device which will improve its performance.

 

          (b)  An equipment manufacturer who resubmits a preliminary breath test device for reconsideration shall provide a reason for resubmission in writing or show that specific improvements to the device will meet the criteria for approval in accordance with Saf-C 6802.01.

 

          (c)  The laboratory shall conduct an evaluation of a preliminary breath test device resubmitted for approval in accordance with Saf-C 6802.03 and report its findings in accordance with Saf-C 6802.04.

 

Source.  (See Revision Note at Chapter Heading for Saf-C 6800) #9685, eff 3-23-10; ss by #12518, eff 4-23-18

 


 

APPENDIX

 

Rule

Specific State Statute the Rule Implements

 

 

Saf-C 6801

RSA 541-A:7

Saf-C 6802

RSA 265-A:15, II and III